LPTH-1
Site-Log Thermistor Datalogger
|
|
Description |
|
|
|
The
Site-Log LPTH-1 is an 7-channel, battery powered, stand alone
thermistor data logger. The logger records up to 4Mb of data and
stores it in non-volatile flash memory for later retrieval. Featuring
an aluminum enclosure and conformal coating PCB, the Site-Log data
logger has excellent performance in the harshest industrial
environment. Featuring a 16-bit ADC and programmable input range, the
Site-Log data loggers are well suited to science and laboratory
applications where precise and accurate measurement is critical.
|
|
Features |
|
|
|

LPTH-1
Supplied
with USB lead
SiteView
software purchased separately
>>
User
Manual >> |
|
|
Description |
|
|
The
Site-Log datalogger range are portable battery powered data loggers
which allow recording information and saving measurements to a 4MB
flash memory for later retrieval. The internal lithium battery
provides up to 10 years of instantaneous logging operation when
sampling at interval of one minute. Site-Log data loggers features a
wide sampling interval range from 20 milliseconds to 12 hours,
providing alternative to data acquisition devices and paper chart recorders.
|
Sampling
Interval |
Enabled
Channel |
Logging
Capacity |
|
1
minute |
1 |
3.98
years |
|
1
minute |
2 |
727
days |
|
1
minute |
8 |
181
days |
|
10
seconds |
1 |
242
days |
|
10
seconds |
2 |
121
days |
|
10
seconds |
8 |
30
days |
|
1
second |
1 |
24
days |
|
1
second |
2 |
12
days |
|
1
second |
8 |
3
days |
|
100
milliseconds |
1 |
58
hours |
|
100
milliseconds |
2 |
29
hours |
|
100
milliseconds |
8 |
7.2
hours |
Site-Log
data logger contains one internal thermistor channel, and seven
external process signal channels. The 16-bit analog-to-digital
converter makes the measurements more precise and accurate. Besides
the on-board thermistor channel, the LPTH-1 logger has seven external
thermistor input channels used to measure temperature.
The
following figure illustrates the correct input connections:

The
Site-Log data logger also provides alarm outputs controlled by the
low and high programmable alarm thresholds of each enabled channel.
The alarm control outputs can be used to turn on/off external alarm
devices such as strobe lights, sirens or alarm telephone diallers.

Site-Log
data logger also has an excitation control output used to switch the
power supply of the external transducer/transmitter, so that these
are only powered during a reading, saving on transducer battery consumption.

Site-Log
data logger also has a rugged, splash resistant aluminum enclosure
and internal electronics are protected against moisture ingress by
conformal coating.
All
Site-Log series data loggers come with an USB and an AUX ports. The
USB port is for connecting directly to the computer. The AUX port is
for connecting to a serial communication device like a Modem or an
Serial Device Server when the logger needs to be accessed remotely.

|
|
SiteView
Software |
|
SiteView
is a PC based application which works with Site-Log Series data
loggers for downloading, configuration and data analysing and
plotting. Its user-friendly graphic interface plus powerful
functionalities fit both novice and advanced users.
The
versatility of custom equation and custom-line equation handle
complicated measurement requirements.
|
|

SSV-1
|
SiteView
Manual >> |
|
Features
Logger
Interaction
Graphing
Measurement
conversion
Data
Importing/Exporting
-
SiteView
File:
- Equations
& Units
- Import,
Export Equation files and Unit files.

|
|
Ethernet
and Modem Support |
|
|
|
|
There
is an AUX port for connecting to a serial communication device like
a Modem or an Serial Device Server when the logger needs to be
accessed remotely. A SL-Rs232 converter cable is available that
terminates in a standard RS232 connector.
|

SL-RS232
|
SL-RS232
Converter is used to convert Site-Log data logger's TTL level
communication signal to RS232 standard.
Applications
of SL-RS232 Converter:
|
|

|
RS232
to
Ethernet Converters
A
range of general purpose RS232 to Ethernet or WiFi converters to
allow the Site-Log range to be configured and read over a LAN or WiFi
Network (requires the SL-RS232)
More
Information >>
 |
|
 |
WirelessCOM/G10
GPRS Modem
The
WirelessCOM/G10 will allow the
Site-Log range to be configured and read remotely over a GPRS mobile
phone network (requires the SL-RS232 and a suitable GPRS SIM Card))
More
Information >>
|
|
|
|
|
|
|
Phone
+44(0)1765 677425 | Fax +44(0)1765 676650 |