Home Page

Contact Us

My Basket

Checkout

LogBox Universal Input Dataloggers


Description

 
A range of low-cost self-contained weatherproof stand alone dataloggers which accurately record data in non-volatile memory. The LogBox-AA datalogger has 2 universal analogue input channels whilst the LogBox-DA and a sinlge universal analogue and a single pulse counter input. Setup and data retrieval is done in a PC via an infrared interface through the use of LOG CHART, a Windows compatible software which plots and prints graphs, lists loggings and exports data to spreadsheets. Special mathematical functions can be programmed. Data can also be retrieved using a Palm PDA.

 

Features

 

 

 LogBox-AA 2 x Universal Analogue Inputs

AA User Manual >>

 

LogBox-DA 1 x Pulse, 1 x Analogue Inputs

DA User Manual >>

 

RS232 IR Interface

USB IR Interface

 

 

LogBox-AA

  • 2 universal multi-sensor channels.

  • Accepts t/c J,K,T,E,N,R,S,B; Pt100, 0-10V, 0-50mV 4-20mA
  • 12 bit resolution, 4096 levels.
  • Accuracy: 0,15% of full scale.

LogBox-DA

  • Channel 1: NPN, PNP, Pulse or Dry Contact

  • Max. frequency: 4000 Hz or 20 Hz
  • Maximum counting: 65,535 per period.
  • Channel 2: 0-20mA, 4-20mA, 0-50mV and 0-10Vdc

Common Features

  • Capacity: 16000 measurements per channel

  • Recording rate: from 10 seconds to 18 Hours
  • Can log maximum/minimum/average values
  • Logging started via front panel button, external contact closure, preser date/time or Palm PDA command
  • Powered by internal 3.6V lithium battery.
  • Battery life: typically 1 year
  • Built-in sensor PSU switch - automatically swicthes on power to sensors when taking a reading, then switches power off
  • Operating temperature -40°C to 75°C.
  • IP65 ABS enclosure
  • Dimensions: 64x58x35mm
  • Supplied with LogChart II Config and Charting Software
  • Supplied with LogChart PalmOS software for Palm PDA users

 

Description

 

The LogBox-AA self-contained logger has up to 2 universal analogue input channels and accurately records the measurements in non-volatile memory. The input channels can be Pt100, Thermocouple (J, K, T, E, N, R, S or B), Voltage (0 to 50 mV or 0 to 10 V) or current (0 to 20mA or 4 to 20mA). Sensors are connected to internal screw terminals via logbox cable glands.

The LogBox-DA is similar to the -AA but has a single pulse counter and a single analogue input. Suitable for connection to flow meters, gas/electricity/water metes and other pulse-output devices.

The LogBox is powered fom an internal 3.6V battery.

Setup and data retrieval is done in a PC via an infrared interface through the use of LogChart-II, a Windows compatible software package.

In addition to LogBox configuration, LogChart-II can be used to retrieve data and for data presentation in table or chart formats. Data can be printed or exported to Spreadsheets such as Excel. Special mathematical functions can be programmed.

Some novel features of the LogBox include;

  • Data capture modes - Data can be acquired via different capture modes:

    • Single - the logger records the instantaneous value at each log rate interval

    • Mean - the mean of ten measurements within the time interval is storedm
    • Minimum or maximum - the minimum or maximum values are stored

  • The LogBox-AA has a built-in power switch which automatically swicthes on external power to sensors when taking a reading, then switches power off again, thus extending external sensor battery life.

  • Data Logging can be started or stopped in various ways;

    • LogBox pushbutton

    • External contact closure
    • Pre-set date and time
    • Immediately
    • Palm PDA Command

 

LogChart Software

 

The LogBox is configured using the included LogChart software. Data is transmitted to the LogBox through an Infrared wand connected to the PC's RS232 or USB port. The Parameter Configuration window in LogChart lets you configure the following parameters:

  • Logging session title

  • Channel setup - name, type, limits, scaling
  • start log options - immediatley or timed
  • log rate
  • logging options - stop when full or roll-over

After the logging session is complete, data can then be downloaded again by LogChart and displayed in table and chart formats. Data files can then be saved as ASCII text files for importation into common Spreadsheets. It is even to monitor data in real-time.

 

Palm PDA Users

 

Most of the functionality of the LogChart II is available for the PDA Palm running the LogChart PalmOS software. The program is supplied with the logger. The program is installed on the Palmtop through a HotSync process (data synchronization between a Palmtop and a PC).

Starting the LogChart PalmOS application will display the Recorded Data screen on the Palm where it is possible to access the LogBox configuration to change settings and collect data, as well as to access collected data.

HotSync of data stored in a PDA to a PC is performed through a conduit installed together with the LogChart Palm-OS. The conduit converts the data collected by the LogChart Palm-OS to a file compatible with the LogChart-II software.

 

 

Specifications

 

Input Type

LogBox-AA

  • Universal input that allows direct sensor measuring: thermocouples J, K, T, N, E, R, S and B, Pt100 RTDs, 0-10V, 0-50mV and 4-20mA

LogBox-DA

  • Channel 1: NPN, PNP, Pulse or Dry Contact

  • Max. frequency: 4000 Hz or 20 Hz
  • Maximum counting: 65,535 per period.
  • Channel 2: 0-20mA, 4-20mA, 0-50mV and 0-10Vdc

ADC Resolution

12-bit (4096 levels)

Accuracy

  • Thermocouples J, K, T e N: 0.25% of full scale ± 1oC

  • Thermocouples E, R, S e B: 0.25% of full scale ± 3oC
  • Pt100: 0.15% of full scale
  • Linear current or voltage: 0.15% of full scale

PT100 Excitation Current

250µA

Input Impedance

  • 100ohm for 4-20mA inputs

  • 10Mohm for thermocouples and milivolts.

Storage Capacity

16000 readings per channel

Data off-load time

8 seconds

Log rate range

10 secs to 18 Hours

Log Start options

  • Immediately

  • Button Press
  • Date/Time
  • external contact closure
  • Palm command

Recording Mode

  • Instant value

  • Average of up to 10 readings for each recording
  • Minimum or maximum value of up to 10 readings for each recording

Logging end

  • At full memory

  • After a preset record number
  • Never stops: overwrite oldest records

Software platform

LOGCHART, Windows 95,98, NT, XP compatible. It configures, reads and displays data on screen in many ways (list, graph, text files to be exported to other software). It accepts plots overlay

Power

3.6V lithium battery (1/2 AA)

Battery life (estimated)

Typically 1 year (1 download daily, 5min log rate)

Working temperature

From -40°C to 70°C (thermocouple cold junction compensation is made between 0°C and 70°C)

Dimensions

60mm x 70mm x 35mm

 

 

Phone +44(0)115 925 8412 | Fax +44(0)115 925 9757