LFC-1
Site-Log Datalogger - 4-20mA
|
|
Description |
|
|
|
The
Site-Log LPVB-1 is an 7-channel, battery powered, stand alone 4-20mA
current data logger. The logger records up to 4Mb of data and stores
it in non-volatile flash memory for later retrieval. Input current
signals can be from sensors, transducers, transmitters or any other
common voltage sources.
Featuring
an aluminum enclosure and conformal coating PCB, the Site-Log data
logger has excellent performance in the harshest industrial
environment. Featuring a 16-bit ADC and programmable input range, the
Site-Log data loggers are well suited to science and laboratory
applications where precise and accurate measurement is critical.
|
|
Features |
|
|
|

LFC-1
Supplied
with USB lead
SiteView
software purchased separately
>>
User
Manual >>
|
|
|
Description |
|
|
The
Site-Log datalogger range are portable battery powered data loggers
which allow recording information and saving measurements to a 4MB
flash memory for later retrieval. The internal lithium battery
provides up to 10 years of instantaneous logging operation when
sampling at interval of one minute. Site-Log data loggers features a
wide sampling interval range from 20 milliseconds to 12 hours,
providing alternative to data acquisition devices and paper chart recorders.
|
Sampling
Interval |
Enabled
Channel |
Logging
Capacity |
|
1
minute |
1 |
3.98
years |
|
1
minute |
2 |
727
days |
|
1
minute |
8 |
181
days |
|
10
seconds |
1 |
242
days |
|
10
seconds |
2 |
121
days |
|
10
seconds |
8 |
30
days |
|
1
second |
1 |
24
days |
|
1
second |
2 |
12
days |
|
1
second |
8 |
3
days |
|
100
milliseconds |
1 |
58
hours |
|
100
milliseconds |
2 |
29
hours |
|
100
milliseconds |
8 |
7.2
hours |
Site-Log
data logger contains one internal thermistor channel, and seven
external process signal channels. The 16-bit analog-to-digital
converter makes the measurements more precise and accurate. Besides
the on-board thermistor channel, the LFC-1 logger has seven external
4-20mA current input channels used to measure typical process control signals.
The
following figure illustrates the correct input connections:

The
Site-Log data logger also provides alarm outputs controlled by the
low and high programmable alarm thresholds of each enabled channel.
The alarm control outputs can be used to turn on/off external alarm
devices such as strobe lights, sirens or alarm telephone diallers.

Site-Log
data logger also has an excitation control output used to switch the
power supply of the external transducer/transmitter, so that these
are only powered during a reading, saving on transducer battery consumption.

Site-Log
data logger also has a rugged, splash resistant aluminum enclosure
and internal electronics are protected against moisture ingress by
conformal coating.
All
Site-Log series data loggers come with an USB and an AUX ports. The
USB port is for connecting directly to the computer. The AUX port is
for connecting to a serial communication device like a Modem or an
Serial Device Server when the logger needs to be accessed remotely.

|
|
SiteView
Software |
|
SiteView
is a PC based application which works with Site-Log Series data
loggers for downloading, configuration and data analysing and
plotting. Its user-friendly graphic interface plus powerful
functionalities fit both novice and advanced users.
The
versatility of custom equation and custom-line equation handle
complicated measurement requirements.
|
|

SSV-1
|
SiteView
Manual >> |
|
Features
Logger
Interaction
Graphing
Measurement
conversion
Data
Importing/Exporting
-
SiteView
File:
- Equations
& Units
- Import,
Export Equation files and Unit files.

|
|
Ethernet
and Modem Support |
|
|
|
|
There
is an AUX port for connecting to a serial communication device like
a Modem or an Serial Device Server when the logger needs to be
accessed remotely. A SL-Rs232 converter cable is available that
terminates in a standard RS232 connector.
|

SL-RS232
|
SL-RS232
Converter is used to convert Site-Log data logger's TTL level
communication signal to RS232 standard.
Applications
of SL-RS232 Converter:
|
|

|
RS232
to
Ethernet Converters
A
range of general purpose RS232 to Ethernet or WiFi converters to
allow the Site-Log range to be configured and read over a LAN or WiFi
Network (requires the SL-RS232)
More
Information >>
 |
|
 |
WirelessCOM/G10
GPRS Modem
The
WirelessCOM/G10 will allow the
Site-Log range to be configured and read remotely over a GPRS mobile
phone network (requires the SL-RS232 and a suitable GPRS SIM Card))
More
Information >>
|
|
|
|
|
|
|
Phone
+44(0)1765 677425 | Fax +44(0)1765 676650 |